top of page
Potable Water Quality Monitoring
Environmental Water Quality Monitoring
Pressure & Transient Pressure Monitoring
Underground Sewer & Trade Waste Management
Wireless Flow Meters
Solar Powered Flow Meters
TracWater Pump & Motor Management System

TracWater TW-4718A Online Optical Dissolved Oxygen Sensor
What's In The Box
-
TW-4718A Dissolved Oxygen Sensor
-
Dissolved Oxygen Sensor Guard
-
10m Cable (default), Customisable
-
User Manual
-
Delivery Inspection Report
Applications
Environmental Monitoring:
Monitor water quality in natural sources like rivers, lakes, oceans, and reservoirs for environmental assessment and conservation.
Aquaculture Industry:
Manage dissolved oxygen levels in aquaculture systems to ensure optimal conditions for aquatic organisms.
Wastewater Management:
Used in wastewater treatment plants to monitor and regulate oxygen levels in the water, aiding in effective wastewater treatment to meet environmental standards.
Scientific Research:
In laboratory settings, dissolved oxygen sensors are crucial for research in biology, chemistry, and environmental science, enabling scientists to study and analyse oxygen levels in various environments.
CATEGORIES
Water Quality Sensors
.png)
.png)
TracWater TW-4718A Online Optical Dissolved Oxygen Sensor
Model number: TW-4718A
Power: 0.1W
Dimension: Φ26mm*187mm (with protective cover)
Customise: OEM, ODM
Product description:
TracWater TW-4718A Online Optical Dissolved Oxygen Sensor is designed for measuring dissolved oxygen concentrations in liquids.
Overview
TracWater TW-4718A Online Optical Dissolved Oxygen sensor incorporates the latest generation of fluorescence lifetime technology. This advanced optical sensor technology provides precise dissolved oxygen (DO) monitoring without signal drift or flow limitations. The design eliminates the need for electrolyte solution, reducing maintenance requirements. Additionally, the sensor is equipped with a built-in temperature sensor that enables automatic temperature compensation to ensure measurement accuracy under varying conditions. The DO sensor outputs a MODBUS/RS-485 digital signal for easy integration.
Features
-
Digital sensor with RS-485 output and MODBUS protocol support.
-
Utilises optical sensor technology for precise dissolved oxygen (DO) monitoring, ensuring no signal drift and eliminating flow limitations.
-
Features excellent anti-interference capabilities, significantly reducing maintenance costs as no electrolyte replacement is required.
-
All calibration data is stored directly in the sensor, enabling on-site installation and immediate use.
Specifications
Name
Online Optical Dissolved Oxygen Sensor
Model
TW-4718A
Measuring principle
Fluorescence method
Range
0-50mg/L or 0-500% Saturation
Accuracy
0 -20mg/L ±1% or ±0.2mg/L
>20mg/L:±5% or ±0.6mg/L
Temperature range
0-50℃
Maximum water depth
60m
IP range
IP68
Interface
RS-485,MODBUS protocol
Power supply
0.1W (DC 6-24V, >50mA)
Installation
G3/4 thread, immersion installation
Size
Φ26mm*187mm
Cable length
10 m (default), customisable
Calibration
1-point or 2-point calibration
Sensor housing material
Ti
bottom of page